The Premier Combo Plus TENS/EMS is a fully digital battery operated pulse generator that sends electrical impulses to the nerves and underlying muscle groups. This unit is a combination stimulator of TENS and EMS which can be used for pain relief and muscle stimulation. The device is provided with two controllable output channels, each independent of the other. A pair of electrodes can be connected to each output channel. The intensity level and settings are controlled by press buttons.

 

MECHANISM

TECHICAL DESCRIPTION

01

Channel

Dual, isolated between channels

02

Pulse Amplitude
Adjustable, 0-110 mA peak into 500 ohm load each channel.

03

Wave Form
Asymmetrical Bi-Phasic Square Pulse

04

Voltage
0 to 55V (Load: 500 ohm)

05

Power source
One 9 Volt Battery. Charger Optional not included 

06

Size
11.8cm(L) x 6 cm(W) x 3.1cm(H)

07

Weight
157 grams with battery.

08

Pulse Rate
Adjustable, from 2 to 150 Hz, 1 Hz/step

09

Pulse Width
Adjustable, from 50 to 300 microseconds, 10 µs/step

10

On Time
Adjustable, 2~90 seconds , 1 Sec./ step

11

Off Time
Adjustable, 2~90 seconds , 1 Sec./ step

12

Ramp Time
Adjustable, 1~8 seconds, 1 Sec./ step, The "On"time will increase and decrease in the setting value.

13

Mode
Six TENS Modes: B(Burst), N(Normal),M (Modulation Rate & Width),S1( Modulation Width), S2 (Modulation Width) and P Three EMS Modes:C(Constant), S (Synchronous), A(Alternate)

14

Burst Mode
Burst rate: Adjustable, 0.5 ~ 5Hz
Pulse width adjustable, 50~300 µs
Frequency fixed = 100 Hz

15

Normal Mode
The pulse rate and pulse width are adjustable. It generates continuous stimulation based on the setting value.

16

Modulation Mode
Modulation mode is a combination of pulse rate and pulse width modulation. The pulse rate and width are automatically varied in a cycle pattern. The pulse    width  is decreased by 50% from its original setting        in 0.5 second, then the pulse rate is decreased by 50% from  its original setting in 0.5 second. Total cycle time   is 1 second. In this mode, pulse rate(2-150Hz) and pulse width(50-300µs) are fully adjustable.

17

S1 Mode
Pulse width is automatically varied in a cyclic pattern over a nominal 10 second period. Pulse width  decreases over a period of 4 seconds from the initial setting to a value 40% less.
The narrower pulse width continues for 1 second.It    then increases over a period of 4 seconds to its initial   setting. The cycle is then repeated. Pulse rate and  pulse width are fully adjustable
18
S2 Mode
Pulse width is automatically varied in a cyclic pattern over a nominal 10 second period. Pulse width  decreases over a period of 4 seconds from the initial setting to a value 70% less.
The narrower pulse width continues for 1 second. It   then increases over a period of 4 seconds to its initial setting. The cycle is then repeated. Pulse rate and   pulse width are fully adjustable.
19
Constant Mode(C)
The pulse rate and pulse width are adjustable. It generates continuous stimulation is delivered .
20
Synchronous Mode(S)
Output from both channels occurs synchronously. The "ON" time includes "Ramp Up" and "Ramp Down" time. Therefore, the setting of ON Time should be no less  than two times of the "Ramp" time in this mode.
21
Alternate Mode(A)
The stimulation of the CH2 will occur after Mode(A) the 1st contraction of CH1 is completed. In this mode, the setting of ON Time should be no less than two times      of the "Ramp" time. The OFF Time should be equal        t o or greater than the ON Time.
ON TIME ≥ Ramp up + Ramp down
OFF TIME ≥ IN TIME
22
Mode P

The pre-set parameters of the 9 programs are as given below:

Program
Mode
Pulse Rate
Pulse Width
Timer
P1
Constant
80Hz
180µs
Continue
P2
Burst
100Hz
(Burst Rate:2Hz)
180µs
Continue
P3
-70% P.W.Modulation
80Hz
50µs-180µs
Continue
P4
Mixed Frequency
15Hz in 3 Sec 
/2Hz in 3 Sec
180µs
Continue
P5
Mixed Frequency
80Hz in 3 Sec /
2Hz in 3 Sec
180µs
Continue
P6
Constant
10Hz
180µs
Continue
P7
Constant
80Hz
60µs
Continue
P8
Constant
80Hz
180µs
30 Minutes
P9
Burst
100Hz(Burst Rate:2Hz)
180µs
30 Minutes
23
Timer
Adjustable, from 5 to 60 minutes minutes and continue(C), 5 minutes each step.
24
Patient 
Compliance Meter
This unit can store 60 sets of operation records. Total recorded time is 999 hours.
25
Low Battery Indicator
A low battery indicator will show up when the battery is low.
26
Operating Condition
Temperature:0°~40°C
Relative Humidity: 30%~75%
Atmosphere Pressure : 700Hpa~1060Hpa
27
Remark
There may be up to a +/-10% tolerance of all  parameters and +/-20% tolerance of output            amplitude & voltage.