Premier EMS Plus EM-6200A Everyway Taiwan
Regular price
Rs.10,800.00
Sale
The Premier EMS is a fully digital battery operated pulse generator that sends electrical impulses to the nerves and underlying muscle groups. This unit is EMS which can be used for muscle stimulation. The device is provided with two controllable output channels, each independent of the other. A pair of electrodes can be connected to each output channel. The intensity level and settings are controlled by press buttons.
|
MECHANISM |
TECHICAL DESCRIPTION |
01 |
Channel |
Dual, isolated between channels |
02 |
Pulse Amplitude
|
Adjustable, 0-110 mA peak into 500 ohm load each channel. |
03 |
Wave Form
|
Asymmetrical Bi-Phasic Square Pulse |
04 |
Voltage
|
0 to 55V (Load: 500 ohm) |
05 |
Power source
|
One 9 Volt Battery. |
06 |
Size
|
11.8cm(L) x 6 cm(W) x 3.1cm(H) |
07 |
Weight
|
157 grams with battery. |
08 |
Pulse Rate
|
Adjustable, from 2 to 150 Hz, 1 Hz/step |
09 |
Pulse Width
|
Adjustable, from 50 to 300 microseconds, 10 µs/step |
10 |
On Time
|
Adjustable, 2~90 seconds , 1 Sec./ step |
11 |
Off Time
|
Adjustable, 2~90 seconds , 1 Sec./ step |
12 |
Ramp Time
|
Adjustable, 1~8 seconds, 1 Sec./ step, The "On"time will increase and decrease in the setting value. |
13 |
Mode
|
Three EMS Modes:C(Constant), S (Synchronous), A(Alternate) |
14
|
Constant Mode(C)
|
The pulse rate and pulse width are adjustable. It generates continuous stimulation is delivered . |
15
|
Synchronous Mode(S)
|
Output from both channels occurs synchronously. The "ON" time includes "Ramp Up" and "Ramp Down" time. Therefore, the setting of ON Time should be no less than two times of the "Ramp" time in this mode. |
16
|
Alternate Mode(A)
|
The stimulation of the CH2 will occur after Mode(A) the 1st contraction of CH1 is completed. In this mode, the setting of ON Time should be no less than two times of the "Ramp" time. The OFF Time should be equal to or greater than the ON Time. ON TIME≥ Ramp up + Ramp down OFF TIME≥IN TIME |
17
|
Timer
|
Adjustable, from 5 to 60 minutes minutes and continue(C), 5 minutes each step. |
18
|
Patient
Compliance Meter |
This unit can store 60 sets of operation records. Total recorded time is 999 hours. |
19
|
Low Battery Indicator
|
A low battery indicator will show up when the battery is low. |
20
|
Operating Condition
|
Temperature:0°~40°C Relative Humidity: 30%~75% Atmosphere Pressure : 700Hpa~1060Hpa |
21
|
Remark
|
There may be up to a +/-10% tolerance of all parameters and +/-20% tolerance of output amplitude & voltage. |