Premier EMS Plus EM-6200A Everyway Taiwan

Premier EMS Plus EM-6200A Everyway Taiwan

Regular price Rs.10,800.00 Sale

The Premier EMS is a fully digital battery operated pulse generator that sends electrical impulses to the nerves and underlying muscle groups. This unit is EMS which can be used for muscle stimulation. The device is provided with two controllable output channels, each independent of the other. A pair of electrodes can be connected to each output channel. The intensity level and settings are controlled by press buttons.

 

MECHANISM

TECHICAL DESCRIPTION

01

Channel

Dual, isolated between channels

02

Pulse Amplitude
Adjustable, 0-110 mA peak into 500 ohm load each channel.

03

Wave Form
Asymmetrical Bi-Phasic Square Pulse

04

Voltage
0 to 55V (Load: 500 ohm)

05

Power source
One 9 Volt Battery.

06

Size
11.8cm(L) x 6 cm(W) x 3.1cm(H)

07

Weight
157 grams with battery.

08

Pulse Rate
Adjustable, from 2 to 150 Hz, 1 Hz/step

09

Pulse Width
Adjustable, from 50 to 300 microseconds, 10 µs/step

10

On Time
Adjustable, 2~90 seconds , 1 Sec./ step

11

Off Time
Adjustable, 2~90 seconds , 1 Sec./ step

12

Ramp Time
Adjustable, 1~8 seconds, 1 Sec./ step, The "On"time will increase and decrease in the setting value.

13

Mode
Three EMS Modes:C(Constant), S (Synchronous), A(Alternate)
14
Constant Mode(C)
The pulse rate and pulse width are adjustable. It generates continuous stimulation is delivered .
15
Synchronous Mode(S)
Output from both channels occurs synchronously. The "ON" time includes "Ramp Up" and "Ramp Down" time. Therefore, the setting of ON Time should be no less than two times of the "Ramp" time in this mode.
16
Alternate Mode(A)
The stimulation of the CH2 will occur after Mode(A) the 1st contraction of CH1 is completed. In this mode, the setting of ON Time should be no less than two times of the "Ramp" time. The OFF Time should be equal to or greater than the ON Time.
ON TIME≥ Ramp up + Ramp down
OFF TIME≥IN TIME
17
Timer
Adjustable, from 5 to 60 minutes minutes and continue(C), 5 minutes each step.
18
Patient 
Compliance Meter
This unit can store 60 sets of operation records. Total recorded time is 999 hours.
19
Low Battery Indicator
A low battery indicator will show up when the battery is low.
20
Operating Condition
Temperature:0°~40°C
Relative Humidity: 30%~75%
Atmosphere Pressure : 700Hpa~1060Hpa
21
Remark
There may be up to a +/-10% tolerance of all parameters and +/-20% tolerance of output amplitude & voltage.