Premier Tens Plus EM-6100A Everyway Taiwan
Regular price
Rs.10,800.00
Sale
The Premier TENS is a fully digital battery operated pulse generator that sends electrical impulses to the nerves and underlying muscle groups. This unit is a TENS used for pain relief . The device is provided with two controllable output channels, each independent of the other. A pair of electrodes can be connected to each output channel. The intensity level and settings are controlled by press buttons.
|
MECHANISM |
TECHICAL DESCRIPTION |
01 |
Channel |
Dual, isolated between channels |
02 |
Pulse Amplitude
|
Adjustable, 0-110 mA peak into 500 ohm load each channel. |
03 |
Wave Form
|
Asymmetrical Bi-Phasic Square Pulse |
04 |
Voltage
|
0 to 55V (Load: 500 ohm) |
05 |
Power source
|
One 9 Volt Battery. |
06 |
Size
|
11.8cm(L) x 6 cm(W) x 3.1cm(H) |
07 |
Weight
|
157 grams with battery. |
08 |
Pulse Rate
|
Adjustable, from 2 to 150 Hz, 1 Hz/step |
09 |
Pulse Width
|
Adjustable, from 50 to 300 microseconds, 10 µs/step |
10 |
Mode
|
Five TENS Modes: B(Burst), N(Normal),M (Modulation Rate & Width),S1( Modulation Width), S2 (Modulation Width) |
11 |
Burst Mode
|
Burst rate: Adjustable, 0.5 ~ 5Hz Pulse width adjustable, 50~300 µs Frequency fixed = 100 Hz |
12 |
Normal Mode
|
The pulse rate and pulse width are adjustable. It generates continuous stimulation based on the setting value. |
13 |
Modulation Mode
|
Modulation mode is a combination of pulse rate and pulse width modulation. The pulse rate and width are automatically varied in a cycle pattern. The pulse width is decreased by 50% from its original setting in 0.5 second, then the pulse rate is decreased by 50% from its original setting in 0.5 second. Total cycle time is 1 second. In this mode, pulse rate(2-150Hz) and pulse width(50-300µs) are fully adjustable. |
14 |
S1 Mode
|
Pulse width is automatically varied in a cyclic pattern over a nominal 10 second period. Pulse width decreases over a period of 4 seconds from the initial setting to a value 40% less. The narrower pulse width continues for 1 second. It then increases over a period of 4 seconds to its initial setting. The cycle is then repeated. Pulse rate and pulse width are fully adjustable. |
15
|
S2 Mode
|
Pulse width is automatically varied in a cyclic pattern over a nominal 10 second period. Pulse width decreases over a period of 4 seconds from the initial setting to a value 70% less. The narrower pulse width continues for 1 second. It then increases over a period of 4 seconds to its initial setting. The cycle is then repeated. Pulse rate and pulse width are fully adjustable. |
16
|
Timer
|
Adjustable, from 5 to 60 minutes minutes and continue(C), 5 minutes each step. |
17
|
Patient
Compliance Meter |
This unit can store 60 sets of operation records. Total recorded time is 999 hours. |
18
|
Low Battery Indicator
|
A low battery indicator will show up when the battery is low. |
19
|
Operating Condition
|
Temperature:0°~40°C Relative Humidity: 30%~75% Atmosphere Pressure : 700Hpa~1060Hpa |
20
|
Remark
|
There may be up to a +/-10% tolerance of all parameters and +/-20% tolerance of output amplitude & voltage. |